Everything You Need to Know About Calibrating Thermometers

Views: 7896

Calibration of temperature equipment is when a certain device, sensor or sensor has a change in temperature, power, and physical pressure. And currently there are many calibration equipment for temperature equipment such as: sensor thermometer, hygrometer, electronic thermometer, temperature recorder, temperature recorder, soldering iron thermometer, machine dryers, temperature controllers, thermometers, RTDs, thermocouples, thermistors, bi-metal thermometers, glass thermometers for calibration… they are currently distributed by very famous brands. For example, such as AOIP 1200 Basic, Isotech QuickCal 550, Ametek ITC 155/320/65, Isotech Pegasus 1200, Isotech Gemini LRI Dry Block, Ametek Jofra ATC 650B, Fluke Calibration Model 9150 Thermocouple Furnace, Fluke Calibration 9190A Field Metrology Well…

 

Calibration procedure of temperature equipment

Calibration procedures for temperature equipment should meet the following standards. Ambient temperature from 18-25 degrees Celsius, air humidity less than 80% RH, calibration room must be ventilated, dust-free, not heated from one side, avoid shocks and collisions, recommended to use laboratory - which is strictly controlled on factors such as humidity, temperature ... before and after calibration, avoiding the impact of the environment.

Normally, the calibration procedure of the temperature device will proceed as follows. Put enough insert blocks in the heat bath, ceramic insulation block put at the bottom of the heat bath, then put the metal insert block for the temperature sensors in it, put the ceramic insulation block on the top. And then you plug the temperature sensor to be calibrated into the insert that has been placed in the heat bath. And connect the output of the temperature sensor to a standard display, to read the results returned by the sensor.

Calibrate the temperature device, next you will select the temperature unit (Celsius or Fahrenheit), set the desired temperature value using the up or down arrow operation to increase or decrease the temperature (heating or heating mode). cooling). For example, we set the standard temperature to 35 degrees Celsius, then wait a moment for the temperature to stabilize and then read the result on the screen. If the value is 36 degrees Celsius, there is a difference of 1 degree Celsius, depending on different calibration procedures, we can accept that level of error or not.

Calibration of temperature equipment for sensors, thermal and electrical equipment... in the manufacturing plant is a necessity. Through the calibration process, we can know the working capacity and life of the equipment, thereby deciding whether to continue using or replace the machine to ensure the quality of manufactured products.